User’s Guide
32 Watt Temperature Control Module
TCM-39032
ILX Lightwave Corporation P. O. Box 6310 Bozeman, MT, U.S.A. 59771 :· · · 1-800-459-9459 · ·
www.ilxlightwave.com
U.S. & Canada International Inquiries: 406-586-1244 Fax 406-586-9405
E-mail: support@ilxlightwave.com
70011703_R00_06/03
Przeglądanie stron 0
1 2 3 4 5 6 ... 31 32

Podsumowanie treści

Strona 1 - User’s Guide

User’s Guide32 Watt Temperature Control ModuleTCM-39032ILX Lightwave Corporation P. O. Box 6310 Bozeman, MT, U.S.A. 59771 :· · · 1-800-459-9459 ·

Strona 2

WARRANTYviii TCM-39032 in a container with at least 3 inches (7.5 cm) of compressible packaging material on all sides.We look forward to serving you

Strona 3 - TABLE OF CONTENTS

TCM-39032 1CHAPTER 1                 INTRODUC

Strona 4

INTRODUCTION AND SPECIFICATIONSSpecifications12 TCM-39032CHAPTER 1 Specifications11. All values relate to a one-hour warm-up period.2. Output curre

Strona 5 - General Safety Considerations

INTRODUCTION AND SPECIFICATIONSSpecifications106_03 TCM-39032 3 CHAPTER 15. Temperature control range depends primarily on the type of thermistor a

Strona 6 - SAFETY SYMBOLS

INTRODUCTION AND SPECIFICATIONSInstallation4 TCM-39032CHAPTER 1 InstallationThis section describes the procedures for installing and removing a TCM

Strona 7 - Returning an Instrument

TCM-39032 5CHAPTER 2                 OPERATIO

Strona 8 - Claims for Shipping Damage

OPERATIONSENSOR SELECT Switches6 TCM-39032CHAPTER 2 SENSOR SELECT SwitchesThe SENSOR SELECT and THERM RANGE switches are used to select sensor type

Strona 9 - 06_03 TCM-39032 vii

OPERATIONTEC Connector06_03 TCM-39032 7 CHAPTER 2TEC ConnectorAt the right of center, on the TCM-39032 back panel, the user will find the 15-pin D-

Strona 10 - TCM-39032

OPERATIONBooster Operation8 TCM-39032CHAPTER 2 present, the TEC Interlock Error condition / event will be reported in the TEC Event Status REgister

Strona 11 - Product Overview

OPERATIONBooster Operation06_03 TCM-39032 9 CHAPTER 2

Strona 13 - 06_03 TCM-39032 3

OPERATIONBooster Operation10 TCM-39032CHAPTER 2 

Strona 14 - Installation

TCM-39032 9CHAPTER 3                 CALIBRAT

Strona 15 - OPERATION

CALIBRATIONEnvironmental Conditions10 TCM-39032CHAPTER 3 Environmental ConditionsCalibrate this instrument under laboratory conditions. ILX recomme

Strona 16 - SENSOR SELECT Switches

CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 11 CHAPTER 35 Press and hold the (PARAMETER) SET switch and turn the ADJUST knob until the TEC

Strona 17 - TEC Interlock

CALIBRATIONTEC Calibration Procedures12 TCM-39032CHAPTER 3 the GPIB hardware. Refer to the GPIB manual for details.7 Switch the THERM RANGE switch

Strona 18 - Booster Operation

CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 13 CHAPTER 3If this value is to be measured and entered remotely via a GPIB controlled DMM, fo

Strona 19

CALIBRATIONTEC Calibration Procedures14 TCM-39032CHAPTER 3 Remote Operation LM335 Sensor The following procedure is for calibrating the LM335 senso

Strona 20

CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 15 CHAPTER 3Local Operation RTD Sensor Calibration (with TSC-595 Option)The following procedur

Strona 21 - Recommended Equipment

CALIBRATIONTEC Calibration Procedures16 TCM-39032CHAPTER 3 Remote Operation RTD Sensor Calibration (with TSC-595 Option)The following procedure is

Strona 22 - TEC Calibration Procedures

CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 17 CHAPTER 3Local Operation ITE Current CalibrationThe following procedure is for calibrating

Strona 23 - CHAPTER 3

TABLE OF CONTENTS4_03 TCM-39032 i                

Strona 24

CALIBRATIONTEC Calibration Procedures18 TCM-39032CHAPTER 3 Remote Operation ITE Current CalibrationThe following procedure is for calibrating the I

Strona 25

CALIBRATIONTEC Calibration Procedures06_03 TCM-39032 19 CHAPTER 3After the last “TEC:ITE” value is entered, the “*OPC?” query may be used to determ

Strona 26

CALIBRATIONTEC Calibration Procedures20 TCM-39032CHAPTER 3 

Strona 27

TABLE OF CONTENTSii TCM-39032Remote Operation Thermistor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 11Local Operation AD590 S

Strona 28

TCM-39032 iiiSAFETY AND WARRANTY INFORMATIONThe Safety and Warranty Information section provides details about cautionary symbols used in the manual

Strona 29

SAFETY SYMBOLSiv TCM-39032 SAFETY SYMBOLSThis section describes the safety symbols and classifications. Technical specifications including electrica

Strona 30

WARRANTY06_03 TCM-39032 v WARRANTYILX LIGHTWAVE CORPORATION warrants this instrument to be free from defects in material and workmanship for a perio

Strona 31

WARRANTYvi TCM-39032 Claims for Shipping DamageWhen you receive the instrument, inspect it immediately for any damage or shortages on the packing li

Strona 32

WARRANTY06_03 TCM-39032 vii Comments, Suggestions, and ProblemsTo ensure that you get the most out of your ILX Lightwave product, we ask that you di